Unité d'Imagerie de l'Appareil Locomoteur, Institut Calot, Berck-sur-mer, France.
Eur Spine J. 1994;3(5):240-5. doi: 10.1007/BF02226572.
Magnetic resonance (MR) imaging is often disturbed after spine surgical procedures with or without an implant. Artifacts are induced by ferromagnetic or nonferromagnetic implants and devices and by small metallic particles left by surgical instruments. All metallic artifacts can affect the quality and usefulness of postoperative spine MR examinations. The physical effects caused by the introduction of metal or other conductive materials into a magnetic field and their consequences are presented. The application to postoperative spine MR examinations and solutions to reduce artifacts are discussed.
脊柱手术无论有无植入物,术后磁共振(MR)成像常受到干扰。铁磁性或非铁磁性植入物及器械以及手术器械遗留的小金属颗粒会诱发伪影。所有金属伪影都会影响术后脊柱MR检查的质量和效用。本文介绍了将金属或其他导电材料引入磁场所产生的物理效应及其后果。还讨论了其在术后脊柱MR检查中的应用以及减少伪影的解决方案。