Suppr超能文献

宫内节育器在磁共振成像检查期间的安全性

Safety of intrauterine contraceptive devices during MR imaging.

作者信息

Hess T, Stepanow B, Knopp M V

机构信息

German Cancer Research Center (dkfz), Heidelberg, Germany.

出版信息

Eur Radiol. 1996;6(1):66-8. doi: 10.1007/BF00619957.

Abstract

Three intrauterine contraceptive devices (IUD) were examined in an experimental in vitro setup to test their MR compatibility: Multiload CU375. Nova T (containing copper and silver), and Gyne T. The devices were fixed in a polyacrylamide gel and exposed to the magnetic field and the radiofrequency (RF) of standard MR sequences (spin-echo, turbo-spin-echo, turboFLASH, and magnetization transfer prepared FLASH). The RF power of a turbo-spin-echo sequence was increased by a factor of 5. Temperatures of the IUDs and the gel were measured during the MR examination. No deflection of the IUDs exposed to the magnetic field of 1.5 Tesla was detected. Under normal imaging conditions no temperature increase could be observed. Using an increased RF power, a maximal temperature rise of 0.4 degree C per examination was seen. No differences between the gel and the IUDs were observed, indicating that there is no specific heating of the tested IUDs.

摘要

在一个体外实验装置中对三种宫内节育器(IUD)进行了检测,以测试它们的磁共振兼容性:Multiload CU375、Nova T(含铜和银)以及Gyne T。这些节育器被固定在聚丙烯酰胺凝胶中,并暴露于标准磁共振序列(自旋回波、快速自旋回波、快速小角度激发以及磁化传递准备快速小角度激发)的磁场和射频(RF)下。快速自旋回波序列的射频功率增加了5倍。在磁共振检查期间测量了节育器和凝胶的温度。未检测到暴露于1.5特斯拉磁场中的节育器有偏移。在正常成像条件下未观察到温度升高。使用增加的射频功率时,每次检查的最大温度升高为0.4摄氏度。未观察到凝胶和节育器之间存在差异,这表明所测试的节育器没有特定的发热现象。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验