多次使用后Hyflex电火花加工机床仪器的表面积和体积变化

Changes in the Surface Area and Volume of Hyflex EDM Instruments after Multiple Uses.

作者信息

Büşra Keskin Neslihan, Özyürek Taha, Uğur Aydin Zeliha, Geneci Ferhat, Ocak Mert, Hamdi Çelik Hakan, Helvacioglu-Yigit Dilek

机构信息

Department of Endodontics, Faculty of Dentistry, Yıldırım Beyazıt University, Ankara, Turkey.

Department of Endodontics, Faculty of Dentistry, Bahçeşehir University, İstanbul, Turkey.

出版信息

Acta Stomatol Croat. 2025 Jun;59(2):114-123. doi: 10.15644/asc59/2/1.

Abstract

OBJECTIVE

The aim of the study was to evaluate the surface areas and volumes of HyFlex EDM (HEDM) nickel-titanium (NiTi) files before and after multiple uses in mandibular molar root canal preparation.

MATERIALS AND METHODS

Twelve HEDM files with sizes of 10/.05 and 25/.08 were evaluated by micro-CT before and after a single use or multiple uses in root canal preparation. The mesiobuccal and mesiolingual root canals of thirty-six human mandibular molar teeth were instrumented using twelve 10/.05 (300 rpm and 1.8 Ncm) and twelve 25/.085 Ncm) HEDM files according to the manufacturer's recommendation. HEDM files were evaluated using micro-CT at four different timelines: 1) Intact file, 2) After instrumentation of one canal, 3) After instrumentation of two canals, and 4) After instrumentation of three canals. Each group's volume and surface area were analyzed by applying Friedman's test. Each group comparison was assessed using the Wilcoxon signed-rank test with Bonferroni correction.

RESULTS

Intact HEDM 10/.05 and HEDM 25/.08 files showed significantly larger volumes and surface areas compared to the used instruments (p < 0.05).

CONCLUSIONS

A significant decrease in the volume and surface area of the NiTi files was observed as the number of uses increased regardless of the file group. These findings emphasize the importance for clinicians to understand the structural changes in HEDM files during repeated use to ensure safe instrument reuse and prevent endodontic treatment failure.

摘要

目的

本研究旨在评估HyFlex EDM(HEDM)镍钛(NiTi)锉在多次用于下颌磨牙根管预备前后的表面积和体积。

材料与方法

对12支尺寸为10/.05和25/.08的HEDM锉在根管预备单次使用或多次使用前后进行显微CT评估。根据制造商的建议,使用12支10/.05(300转/分钟和1.8牛顿厘米)和12支25/.08(5牛顿厘米)的HEDM锉对36颗人下颌磨牙的近中颊根和近中舌根进行根管预备。在四个不同时间点使用显微CT对HEDM锉进行评估:1)完整锉;2)一根根管预备后;3)两根根管预备后;4)三根根管预备后。应用Friedman检验分析每组的体积和表面积。每组比较采用经Bonferroni校正的Wilcoxon符号秩检验进行评估。

结果

与使用后的器械相比,完整的HEDM 10/.05和HEDM 25/.08锉显示出明显更大的体积和表面积(p < 0.05)。

结论

无论锉的组别如何,随着使用次数的增加,观察到NiTi锉的体积和表面积显著减小。这些发现强调了临床医生了解HEDM锉在重复使用过程中的结构变化以确保器械安全再利用并防止牙髓治疗失败的重要性。