Stone R G, Weeks L E, Hajdu M, Stinchfield F E
Clin Orthop Relat Res. 1985 Dec(201):26-31.
Sciatic nerve function was monitored intraoperatively by somatosensory evoked potentials (SEP) in 50 patients during total hip arthroplasty. There were 12 instances of temporary nerve compromise in ten patients: three during acetabular preparation due to retraction, one due to reamer contact, six during femoral reaming due to positioning, and two during trial reduction due to inadequate shortening of the femoral neck. The technique for monitoring, causes, and incidence of nerve compromise during total hip arthroplasty are discussed. SEP monitoring may be especially useful in revision surgery.
在50例全髋关节置换术中,通过体感诱发电位(SEP)对坐骨神经功能进行术中监测。10例患者出现12次短暂性神经损伤:髋臼准备时因牵开导致3次,扩孔钻接触导致1次,股骨扩髓时因体位问题导致6次,试行复位时因股骨颈缩短不足导致2次。本文讨论了全髋关节置换术中神经损伤的监测技术、原因及发生率。SEP监测在翻修手术中可能特别有用。