三种用于淀粉样蛋白PET定量分析的商业软件平台的百分克氏单位值一致性分析

Consistency Analysis of Centiloid Values Across Three Commercial Software Platforms for Amyloid PET Quantification.

作者信息

Yoon Hyukjin, Lee Narae, Um Yoo Hyun, Choi Woo Hee

机构信息

Division of Nuclear Medicine, Department of Radiology, St. Vincent's Hospital, College of Medicine, The Catholic University of Korea, 222 Banpo-daero, Seocho-gu, Seoul 06591, Republic of Korea.

Department of Psychiatry, St. Vincent's Hospital, College of Medicine, The Catholic University of Korea, 222 Banpo-daero, Seocho-gu, Seoul 06591, Republic of Korea.

出版信息

Diagnostics (Basel). 2025 Jun 24;15(13):1599. doi: 10.3390/diagnostics15131599.

Abstract

: This study aimed to evaluate the consistency of Centiloid (CL) values calculated using three commercially available software platforms: BTXBrain (v1.1.2), MIM (v7.3.7), and SCALE PET (v2.0.1). : A total of 239 patients who underwent amyloid PET/CT with either F-18 flutemetamol (FMM) or F-18 florbetaben (FBB) were retrospectively analyzed. CL values were calculated using BTXBrain, MIM, and SCALE PET. Linear regression, Passing-Bablok regression, and Bland-Altman analysis were performed to assess the agreement between CL values. Subgroup analyses were conducted for each radiotracer. CL values were compared according to visual interpretation status. : Strong correlations were observed between CL values derived from the three software platforms (R > 0.95). However, Passing-Bablok regression revealed significant proportional bias, with CL values from BTXBrain being lower than others, and CL values from SCALE PET being higher than others as CL values increased. Bland-Altman plots visualized the proportional bias, particularly between BTXBrain and SCALE PET. Subgroup analyses by radiotracer showed similar results. CL values in visually positive scans were significantly higher than those in visually negative scans across all platforms. : The three commercial software programs demonstrated high consistency in CL quantification. However, a notable systematic bias was observed. Further evaluation of various scanner effects and CL calculation methods is warranted to improve the consistency and reproducibility of CL quantification in clinical practice.

摘要

本研究旨在评估使用三个商用软件平台计算的Centiloid(CL)值的一致性:BTXBrain(v1.1.2)、MIM(v7.3.7)和SCALE PET(v2.0.1)。对总共239例接受F-18氟替美莫(FMM)或F-18氟贝他宾(FBB)淀粉样蛋白PET/CT检查的患者进行回顾性分析。使用BTXBrain、MIM和SCALE PET计算CL值。进行线性回归、Passing-Bablok回归和Bland-Altman分析以评估CL值之间的一致性。对每种放射性示踪剂进行亚组分析。根据视觉解读状态比较CL值。在来自三个软件平台的CL值之间观察到强相关性(R>0.95)。然而,Passing-Bablok回归显示出显著的比例偏差,随着CL值增加,BTXBrain的CL值低于其他软件,而SCALE PET的CL值高于其他软件。Bland-Altman图直观显示了比例偏差,特别是在BTXBrain和SCALE PET之间。按放射性示踪剂进行的亚组分析显示了类似结果。在所有平台上,视觉上阳性扫描的CL值显著高于视觉上阴性扫描的CL值。这三个商业软件程序在CL定量方面表现出高度一致性。然而,观察到了明显的系统偏差。有必要进一步评估各种扫描仪效应和CL计算方法,以提高临床实践中CL定量的一致性和可重复性。