Poirier Jasmine, Allain Guillaume, Bansept Marc-Antoine, Hamel Éric, Sauvageau Dominic, Akitegetse Cléophace
Zilia Inc., Quebec City, Canada.
Department of Chemical and Materials Engineering, University of Alberta, Edmonton, Canada.
Biomed Opt Express. 2025 Jun 16;16(7):2739-2755. doi: 10.1364/BOE.566215. eCollection 2025 Jul 1.
Retinal vascular tortuosity is a biomarker associated with various retinal and systemic diseases. However, its clinical application is hindered by the lack of an objective, standardized definition. With the growing number of automated pipelines for the evaluation of tortuosity from retinal images, the impact of image processing, choice of tortuosity index, and retinal camera imaging parameters have been overlooked. This study assesses the robustness of different tortuosity indices to artificial changes in camera imaging parameters and the choice of centerline extraction algorithm to provide a path toward standardization of tortuosity quantification. Five tortuosity indices were implemented and tested on tortuosity-ordered vessel segments using Spearman's Rank Correlation Coefficients for comparison. Comparable results were obtained for all tested indices. Image magnification was found to have minor or no impact on the most tested indices. A key finding from this study is that the same blood vessel imaged in different regions of the field of view of the same imaging system may differ in measured tortuosity. Different centerline extraction algorithms were also found to have a critical impact on tortuosity quantification. Taken as a whole, this study shows the importance of standardizing the analytical procedures in the evaluation of blood vessel tortuosity.
视网膜血管迂曲是一种与多种视网膜及全身性疾病相关的生物标志物。然而,由于缺乏客观、标准化的定义,其临床应用受到了阻碍。随着用于评估视网膜图像中血管迂曲的自动化流程数量不断增加,图像处理、迂曲指数的选择以及视网膜相机成像参数的影响一直被忽视。本研究评估了不同迂曲指数对相机成像参数人工改变的稳健性以及中心线提取算法的选择,以提供一条实现迂曲量化标准化的途径。使用斯皮尔曼等级相关系数对五个迂曲指数在按迂曲程度排序的血管段上进行了实施和测试,以便进行比较。所有测试指数均获得了可比结果。发现图像放大倍数对大多数测试指数影响较小或无影响。本研究的一个关键发现是,在同一成像系统视野的不同区域成像的同一血管,其测量的迂曲度可能不同。还发现不同的中心线提取算法对迂曲量化有至关重要的影响。总体而言,本研究表明了在评估血管迂曲时标准化分析程序的重要性。