Examples of six oxygen concentrators (DeVO2, Dom 10, Econo 2, Hudson, Permox, and Roomate) were evaluated over a 9-28 day period to determine (1) the oxygen yield (% O2) over the flow range 1-4 l min-1; (2) 90% oxygen rise time (90% RT) from a cold start when they were operated at 2 l min-1; (3) accuracy and readability of the flow device; (4) static outlet pressure; (5) major components comprising the product gas (Hudson only); and (6) general characteristics. At an outlet flow of 2 l min-1 the mean % O2 generated by all models, except the Permox (which was lower, mean (SD) 90.5% (3.1%), were between 94% and 95% with a range of less than +/- 0.5%. The Dom 10, Econo 2, and Hudson consistently generated higher oxygen concentrations than the other models at flow rates greater than 2 l min-1. The 90% RT was less than 10.5 minutes for all models. Deviations of up to 22% were observed between predicted and measured flow rates in all models except the DeVO2, Hudson, and Permox. It was possible to set the orifice type flow devices fitted to the Permox and Roomate between indicated flow settings, resulting in cessation of flow. Spectral analysis of the output of one device showed that argon and oxygen were concentrated to similar extents, indicating that the maximal attainable oxygen yield for a molecular sieve concentrator is about 96%.
对六种制氧机(DeVO2、Dom 10、Econo 2、Hudson、Permox和Roomate)进行了为期9至28天的评估,以确定:(1)在1-4升/分钟的流量范围内的氧气产率(%O2);(2)当以2升/分钟运行时,从冷启动开始的90%氧气上升时间(90%RT);(3)流量装置的准确性和可读性;(4)静态出口压力;(5)构成产物气体的主要成分(仅适用于Hudson);以及(6)一般特性。在2升/分钟的出口流量下,除Permox(较低,平均值(标准差)为90.5%(3.1%))外,所有型号产生的平均%O2在94%至95%之间,范围小于±0.5%。在流量大于2升/分钟时,Dom 10、Econo 2和Hudson始终比其他型号产生更高的氧气浓度。所有型号的90%RT均小于10.5分钟。除DeVO2、Hudson和Permox外,所有型号的预测流量与测量流量之间的偏差高达22%。可以在Permox和Roomate配备的孔板型流量装置的指示流量设置之间进行设置,从而导致流量停止。对一台设备的输出进行光谱分析表明,氩气和氧气的浓缩程度相似,这表明分子筛制氧机可达到的最大氧气产率约为96%。