In vivo validation of damping capacity assessment as a diagnostic tool for peri-implant bone loss.

作者信息

Pyo Se-Wook, Kim Gi Youn, Chang Jae-Seung, Kim Sunjai

机构信息

Department of Prosthodontics, Gangnam Severance Dental Hospital, Yonsei University College of Dentistry, 211 Eonju-ro, Gangnam-gu, Seoul, 06273, Republic of Korea.

出版信息

Sci Rep. 2025 Jul 31;15(1):27984. doi: 10.1038/s41598-025-11144-1.

Abstract

This in vivo animal study aimed to evaluate the diagnostic reliability of a damping capacity assessment by correlating dental implant stability with peri-implant bone loss in a beagle dog model. Thirty-two bone-level implants were immediately placed in four beagle dogs without bone grafting. Implants were divided into platform switching and platform matching groups. Each implant was loaded with a superstructure after a 4-week healing period, followed by 8 weeks of functional loading. Damping capacity assessment and resonance frequency analysis were performed every 4 weeks. Peri-implant bone loss, supporting bone volume, and bone-to-implant contact were assessed using micro-CT and histological analysis. Two implants in the platform switching group failed during the study. Among the remaining 30 implants, final implant stability indices by damping capacity assessment ranged from 44 to 80. Peri-implant bone loss varied from 0.25 mm to 6.96 mm. A strong negative correlation was found between final stability indices and bone loss (r = - 0.745; p ≤ 0.0001). Supporting bone volume and bone-to-implant contact ratio showed significant positive correlations with implant stability. Damping capacity assessment demonstrated diagnostic relevance by reflecting peri-implant bone loss and the degree of osseointegration. However, its ability to detect early bone resorption is uncertain.

摘要

这项体内动物研究旨在通过在比格犬模型中将牙种植体稳定性与种植体周围骨丢失相关联,评估阻尼能力评估的诊断可靠性。在四只未进行骨移植的比格犬中立即植入32颗骨水平种植体。种植体分为平台转换组和平台匹配组。在4周的愈合期后,每个种植体都安装了上部结构,随后进行8周的功能加载。每4周进行一次阻尼能力评估和共振频率分析。使用微型计算机断层扫描和组织学分析评估种植体周围骨丢失、支持骨体积和骨与种植体接触情况。平台转换组中有两颗种植体在研究期间失败。在其余30颗种植体中,通过阻尼能力评估得出的最终种植体稳定性指数范围为44至80。种植体周围骨丢失在0.25毫米至6.96毫米之间变化。最终稳定性指数与骨丢失之间存在强烈的负相关(r = - 0.745;p≤0.0001)。支持骨体积和骨与种植体接触率与种植体稳定性呈显著正相关。阻尼能力评估通过反映种植体周围骨丢失和骨结合程度显示出诊断相关性。然而,其检测早期骨吸收的能力尚不确定。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0b2c/12313939/641c338e59cc/41598_2025_11144_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验