Sung Jiwon, Choi Yeonho, Kim Jina, Kim Jun Won, Kim Jihun
Department of Radiation Oncology, Gangnam Severance Hospital, Yonsei University College of Medicine, Seoul, South Korea.
Sci Rep. 2025 Aug 12;15(1):29515. doi: 10.1038/s41598-025-15107-4.
This study aimed to develop and publicly release an in-house software package that converts the binary file format of 2D cine magnetic resonance (MR) images acquired through the Treatment Session Manager (TSM) of MOSAIQ on the Elekta Unity (Elekta AB, Stockholm, Sweden) into standard readable data formats. The software was developed using MATLAB (MathWorks, Natick, MA, USA) and includes an automatic image-sorting algorithm to classify the images into coronal, sagittal, and axial planes. To verify the geometric accuracy of the converted images, they were acquired from an MRgRT motion management QA phantom, both with and without motion. For the converted images without motion, the geometric size of the phantom was measured and compared with the known values. For the images acquired with motion, the magnitudes measured using the converted cine MR images were compared with the artificially introduced motions. The results showed that the 2D cine MR images in the binary file format were successfully converted into the metadata/DICOM format and accurately classified into different planes. The accuracies of the geometric size and motion in the converted 2D cine MR images were greater than 99.6% and 94.2%, respectively. This software is expected to be useful for 1.5 T MR-linac users in analyzing internal organ motion during radiation treatment.
本研究旨在开发并公开发布一个内部软件包,该软件包可将通过瑞典斯德哥尔摩的医科达公司(Elekta AB)的Elekta Unity上的MOSAIQ治疗会话管理器(TSM)获取的二维电影磁共振(MR)图像的二进制文件格式转换为标准可读数据格式。该软件使用MATLAB(美国马萨诸塞州纳蒂克市的MathWorks公司)开发,包括一种自动图像分类算法,可将图像分类为冠状面、矢状面和横断面。为了验证转换后图像的几何精度,从MRgRT运动管理质量保证体模获取了有运动和无运动情况下的图像。对于无运动的转换后图像,测量体模的几何尺寸并与已知值进行比较。对于有运动情况下获取的图像,将使用转换后的电影MR图像测量的幅度与人工引入的运动进行比较。结果表明,二进制文件格式的二维电影MR图像成功转换为元数据/DICOM格式,并准确分类到不同平面。转换后的二维电影MR图像中几何尺寸和运动的精度分别大于99.6%和94.2%。预计该软件对1.5 T MR直线加速器用户在放射治疗期间分析内部器官运动有用。