Suppr超能文献

Multi-Objective CFD Optimization of an Intermediate Diffuser Stage for PediaFlow Pediatric Ventricular Assist Device.

作者信息

Zhussupbekov Mansur, Wu JingChun, Burgreen Greg W, Stelick Scott, Kim Jeongho, Antaki James F

机构信息

Meinig School of Biomedical Engineering, Cornell University, Ithaca, New York, USA.

Advanced Design Optimization, LLC, Irvine, California, USA.

出版信息

Artif Organs. 2025 Aug 14. doi: 10.1111/aor.70004.

Abstract

BACKGROUND

Computational fluid dynamics (CFD) has become an essential design tool for ventricular assist devices (VADs), where the goal of maximizing performance often conflicts with biocompatibility. This tradeoff becomes even more pronounced in pediatric applications due to the stringent size constraints imposed by the smaller patient population. This study presents an automated CFD-driven shape optimization of a new intermediate diffuser stage for the PediaFlow pediatric VAD, positioned immediately downstream of the impeller to improve pressure recovery.

METHODS

We adopted a multi-objective optimization approach to maximize pressure recovery while minimizing hemolysis. The proposed diffuser stage was isolated from the rest of the flow domain, enabling efficient evaluation of over 450 design variants using Sobol sequence, which yielded a Pareto front of nondominated solutions. The selected best candidate was further refined using a local T-search algorithm. We then incorporated the optimized front diffuser into the full pump for CFD verification and in vitro validation.

RESULTS

We identified critical dependencies where longer blades increased pressure recovery but also hemolysis, while the wrap angle showed a strong parabolic relationship with pressure recovery but a monotonic relationship with hemolysis. Counterintuitively, configurations with fewer blades (2, 3) consistently outperformed those with more blades (4, 5) in both metrics. The optimized two-blade design enabled operation at lower pump speeds (14 000 vs. 16 000 RPM), improving hydraulic efficiency from 26.3% to 32.5% and reducing hemolysis by 31%.

CONCLUSION

This approach demonstrates that multi-objective CFD optimization can systematically explore complex design spaces while balancing competing priorities of performance and hemocompatibility for pediatric VADs.