表征采用不同制造技术的中段尿道吊带的结构特性和孔隙率。

Characterizing the structural properties and porosity of mid-urethral slings with varied manufacturing techniques.

作者信息

Knight Katrina, Meyn Leslie, Moalli Pamela

机构信息

Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA, United States.

Magee-Womens Research Institute, Pittsburgh, PA, United States.

出版信息

Front Bioeng Biotechnol. 2025 Aug 7;13:1543808. doi: 10.3389/fbioe.2025.1543808. eCollection 2025.

Abstract

INTRODUCTION

Implantation of mid-urethral slings (MUSs) is a safe and effective approach for the surgical repair of stress urinary incontinence. However, concerns regarding the deformability of the prototype MUS, mechanical cut Gynecare TVT prompted manufacturers to use techniques like laser cutting, heat-sealing, and the inclusion of an interwoven stabilizing suture to decrease deformation with loading. We hypothesized that a laser cut or heat-sealed MUS would be stiffer but deform less, and experience less permanent elongation as compared to a mechanical cut MUS. Additionally, the inclusion of a stabilizing suture would minimize the loss of porosity.

METHODS

Uniaxial tensile testing to failure and cyclic loading was performed to analyze the structural properties and permanent elongation, respectively, of commercially available MUSs Gynecare TVT (mechanical cut), Gynecare TVT Exact (laser cut), ArcTV (laser cut, with and without the stabilizing suture), and Desara Blue (heat-sealed). A custom Mathematica code was used to quantify the porosity of the MUSs following sequential uniaxial loading from 0 N to 10 N.

RESULTS

Desara Blue was significantly stiffer (p-values < 0.05), elongated less at failure (p = 0.002), and experienced less permanent elongation in response to cyclic loading (p-values = 0.001) relative to Gynecare TVT. Similarly, permanent elongation was significantly less (p-values = 0.004) and the stiffness was higher (p = 0.004) for Gynecare TVT Exact as compared to Gynecare TVT. Very little differences in stiffness and no differences in relative elongation at failure nor permanent elongation were observed between ArcTV (without the interwoven suture) and Gynecare TVT (p-values > 0.05). The porosity of all MUSs significantly decreased with loading (p-values < 0.001); except for ArcTV with the stabilizing suture which showed the least amount of deformation (i.e., percent change in porosity decreasing by only 14%, p < 0.001, at 10 N).

DISCUSSION

Overall, heat-sealing decreased deformability at the cost of markedly increasing device stiffness to a point which likely outweighs benefits, and risks increased complications. Laser cutting had different effects on the behavior of TVT Exact and ArcTV suggesting manufacturer technical differences, but overall reduced deformation without a substantial impact on stiffness. An interwoven stabilizing suture minimized the loss of porosity which translates clinically to less deformation and mechanistically to reduced mesh complications.

摘要

引言

植入尿道中段吊带(MUS)是压力性尿失禁外科修复的一种安全有效的方法。然而,对原型MUS(机械切割的Gynecare TVT)可变形性的担忧促使制造商采用激光切割、热封以及加入交织稳定缝线等技术,以减少加载时的变形。我们假设,与机械切割的MUS相比,激光切割或热封的MUS会更硬但变形更小,且永久伸长更少。此外,加入稳定缝线将使孔隙率损失最小化。

方法

分别对市售的MUS(Gynecare TVT(机械切割)、Gynecare TVT Exact(激光切割)、ArcTV(有或无稳定缝线的激光切割)和Desara Blue(热封))进行单轴拉伸至破坏试验和循环加载试验,以分析其结构性能和永久伸长情况。使用自定义的Mathematica代码对MUS从0 N到10 N连续单轴加载后的孔隙率进行量化。

结果

与Gynecare TVT相比,Desara Blue明显更硬(p值<0.05),破坏时伸长更少(p = 0.002),并且在循环加载时永久伸长更少(p值 = 0.001)。同样,与Gynecare TVT相比,Gynecare TVT Exact的永久伸长明显更少(p值 = 0.004)且刚度更高(p = 0.004)。在ArcTV(无交织缝线)和Gynecare TVT之间,未观察到刚度有很大差异,破坏时的相对伸长和永久伸长也无差异(p值>0.05)。所有MUS的孔隙率均随加载而显著降低(p值<0.001);除了带有稳定缝线的ArcTV,其变形量最小(即孔隙率变化百分比在10 N时仅降低14%,p < 0.001)。

讨论

总体而言,热封以显著增加器械刚度为代价降低了可变形性,而增加的刚度可能超过了益处,并增加了并发症风险。激光切割对TVT Exact和ArcTV的性能有不同影响,这表明制造商的技术存在差异,但总体上减少了变形且对刚度没有实质性影响。交织稳定缝线使孔隙率损失最小化,这在临床上意味着变形更小,从机制上讲可减少网片并发症。