Shintani H, Ube S
J Chromatogr. 1985 Nov 8;344:145-56. doi: 10.1016/s0378-4347(00)82015-0.
A procedure is described for the analysis of serum cations and uremic toxins by simultaneous conductivity (CD) and ultraviolet (UV) photometric detection and using phosphoric acid as the eluent. In addition, studies were made on the determination of urea in serum using immobilized urease, and on the simultaneous determination of cationic and anionic compounds. The ammonium ion produced from urea by immobilized urease was determined by CD. No significant differences in amounts were found between the present method and conventional colorimetric methods. There was no significant reduction in immobilized urease activity after two months of continuous use. Simultaneous analysis of cationic and anionic compounds in serum was performed by switching valves in order to change the eluents and to change the channel to the analytical column. No marked base-line variation was observed and satisfactory recovery of anions was achieved.
描述了一种通过同时进行电导率(CD)和紫外(UV)光度检测并使用磷酸作为洗脱液来分析血清阳离子和尿毒症毒素的方法。此外,还进行了使用固定化脲酶测定血清中尿素以及同时测定阳离子和阴离子化合物的研究。通过CD测定固定化脲酶由尿素产生的铵离子。本方法与传统比色法在含量上无显著差异。连续使用两个月后,固定化脲酶的活性没有显著降低。通过切换阀门以改变洗脱液并将通道切换至分析柱,对血清中的阳离子和阴离子化合物进行同时分析。未观察到明显的基线变化,阴离子的回收率令人满意。