Bjerkén Anna, Karampiperi Maria, Tomic Hanna, Dustler Magnus, Rääf Christopher, Tingberg Anders, Bernhardsson Christian, Bakic Predrag R
Medical Radiation Physics, Department of Translational Medicine, Lund University, Skåne University Hospital, SE-205 02 Malmö, Sweden.
Radiation Physics, Department of Hematology, Oncology and Radiation Physics, Skåne University Hospital, SE-205 02 Malmö, Sweden.
Radiat Prot Dosimetry. 2025 Aug 26;201(13-14):960-965. doi: 10.1093/rpd/ncaf037.
In this study, NaCl pellets read with optically stimulated luminescence were evaluated for their potential use as a point dosemeter in breast x-ray imaging. Dosimetry with NaCl pellets has previously been applied to various environmental and medical settings. NaCl pellets have potential in clinical breast dosimetry because they can enable multiple point measurements to be simultaneously conducted in a cost- and time-efficient manner. Using two digital mammography imaging systems, the air kerma response of the NaCl pellets for a standard breast setup was investigated. The air kerma response was observed to be linear, and mathematical fits were successfully used to estimate the cumulative incident air kerma during digital breast tomosynthesis. Deviations from a reference digital dosemeter were 6% and 8% for the two mammography systems, respectively. Measurements conducted at different angles of exposure showed that the NaCl pellets had no angular dependency in the range ± 15°. Finally, the uniformity of the beams was confirmed to avoid possible errors due to the uncertainty of the dosemeters' positions in the exposure field.
在本研究中,对用光学激发发光读数的氯化钠颗粒作为乳腺X线成像中的点剂量计的潜在用途进行了评估。此前,氯化钠颗粒剂量测定已应用于各种环境和医疗场景。氯化钠颗粒在临床乳腺剂量测定中具有潜力,因为它们能够以经济高效的方式同时进行多点测量。使用两个数字乳腺摄影成像系统,研究了氯化钠颗粒在标准乳腺设置下的空气比释动能响应。观察到空气比释动能响应呈线性,并且成功地使用数学拟合来估计数字乳腺断层合成期间的累积入射空气比释动能。两个乳腺摄影系统与参考数字剂量计的偏差分别为6%和8%。在不同曝光角度下进行的测量表明,氯化钠颗粒在±15°范围内没有角度依赖性。最后,确认了射束的均匀性,以避免由于剂量计在曝光场中位置的不确定性而可能产生的误差。