Suppr超能文献

评估Gazepoint GP3 HD在评估听力努力方面的有效性:一项瞳孔测量研究。

Evaluating the Validity of Gazepoint GP3 HD in Assessing Listening Effort: A Pupillometry Study.

作者信息

Rahme Mohamed, Parsa Vijay, Farahani Mojgan, Folkeard Paula, Scollie Susan, Johnsrude Ingrid Suzanne

机构信息

National Centre for Audiology, Western University, London, Ontario, Canada.

Communication Sciences and Disorders, Faculty of Health Sciences, Western University, London, Ontario, Canada.

出版信息

Am J Audiol. 2025 Sep 2;34(3):742-753. doi: 10.1044/2025_AJA-24-00190. Epub 2025 Aug 28.

Abstract

PURPOSE

Individuals with hearing loss typically experience greater listening effort, which is the additional recruitment of cognitive/mental resources such as attention and memory to understand speech and can be aversive and tiring. Reducing effort is an important goal of the hearing health care industry. Pupillometry is an objective and increasingly popular measure of listening effort, but gold standard measures of pupil size are expensive and unwieldy. The purpose of this study was to compare a low-cost and portable pupillometry device (Gazepoint GP3 HD) to a more traditional gold standard pupillometry tool (EyeLink 1000) for indexing listening effort via pupil size.

METHOD

Twenty normal-hearing young adults (age range: 18-23 years) were recruited in this study. Participants' pupil size was measured using the Gazepoint and EyeLink pupillometry devices while listening to Hearing in Noise Test sentences in stationary speech-shaped noise at signal-to-noise ratios (SNRs) ranging from -8 to +8 dB.

RESULTS

Participants' word report accuracy increased from approximately 12% to 100% when the SNRs increased from -8 to +8 dB. Peak pupil diameter decreased for both devices and was smaller with the Gazepoint device. Data quality was comparable for the two devices.

CONCLUSION

Gazepoint appeared to be an effective pupillometry device that records pupil dilation across a wide range of SNRs, without interfering with the auditory task.

摘要

目的

听力损失患者通常会经历更大的听力努力,即额外调动认知/心理资源(如注意力和记忆力)来理解言语,这可能会令人反感且疲惫。减少听力努力是听力保健行业的一个重要目标。瞳孔测量法是一种客观且越来越受欢迎的听力努力测量方法,但瞳孔大小的金标准测量方法昂贵且不便使用。本研究的目的是将一种低成本便携式瞳孔测量设备(Gazepoint GP3 HD)与一种更传统的金标准瞳孔测量工具(EyeLink 1000)进行比较,以通过瞳孔大小来衡量听力努力。

方法

本研究招募了20名听力正常的年轻成年人(年龄范围:18 - 23岁)。在信噪比(SNR)从 -8到 +8 dB的稳态言语噪声环境中,让参与者听噪声中的听力测试句子时,使用Gazepoint和EyeLink瞳孔测量设备测量他们的瞳孔大小。

结果

当信噪比从 -8 dB增加到 +8 dB时,参与者的单词报告准确率从约12%提高到了100%。两种设备的瞳孔直径峰值均下降,且Gazepoint设备的瞳孔直径峰值更小。两种设备的数据质量相当。

结论

Gazepoint似乎是一种有效的瞳孔测量设备,能够在广泛的信噪比范围内记录瞳孔扩张情况,且不会干扰听觉任务。