Determination of a safe protocol for using laser ablation with indocyanine green dye in endodontic treatment. In vitro, in vivo and human study.

作者信息

Leonardo Renato de Toledo, Perdomo Mirtha, Perdomo Marcelo Costa, Giménez María Betania Acevedo, Nishiyama Celso Kenji, Orosco Fernando Accorsi, Garcia Arturo Javier Aranda, Wajima Carolina Sayuri, Cantiga-Silva Cristiane, Vasques Ana Maria Veiga, Faria Flávio Duarte, Dezan Júnior Eloi, Cintra Luciano Tavares Angelo

机构信息

São Paulo State University (UNESP), School of Dentistry, Araraquara, SP, Brazil.

Universidad Autónoma del Paraguay, Asunción, Paraguay.

出版信息

Lasers Med Sci. 2025 Sep 2;40(1):350. doi: 10.1007/s10103-025-04589-3.

Abstract

UNLABELLED

This study aimed to determine a safe protocol for using Laser Ablation therapy (LA) with Indocyanine Green (ICG) to treat infected root canals.

METHODS

In vitro study - The root canal of an extracted human lower incisor was prepared to #35/.06 and was filled with 0.05% ICG dye. The λ 810 nm infrared diode laser was activated during the movement of introducing a #20 fiber into the root canals. The maximum temperature reached in the three root thirds was considered for protocols, modifying the power, pulse interval, and pulse duration. In vivo study - Extracted human lower incisor, lower premolar, and upper molar were selected. Their canals were prepared to #35/0.04, and the teeth were inserted into the subcutaneous tissue of Wistar rats. LA with ICG protocol selected in the in vitro study was performed on each root. The root surface temperature was measured in the three thirds. Then, the subcutaneous tissue adjacent to the roots was removed immediately or after 7 days and subjected to histological analysis in H&E stain. Human study - Thirteen lower incisors indicated for extraction from eleven patients were included. The canals of all teeth were prepared and received LA treatment with ICG in a similar way to the in vivo study. After tooth extraction, the apical thirds of the roots were processed for histopathological analysis.

RESULTS

The in vitro study revealed that the increase in power and pulse duration, or the reduction in pulse interval, determined the temperature rise. The protocol with the highest power that showed a temperature increase below 40 °C was the 2.5 W (power), 300 ms (interval), and 100 ms (duration). The in vivo study revealed that the greatest temperature variation was observed in the lower incisor; for this reason, it was selected for the human study. In the human study, a structure of the periodontal ligament with histological and morphological characteristics comparable to those of healthy tissues and similar to the roots of control teeth was observed in the treated teeth (p < 0.05). No harmful effects were observed on the cementum structure, periodontal ligament, and alveolar bone in any teeth evaluated.

CONCLUSIONS

It was concluded that the LA protocol used is safe and can be performed during the treatment of infected root canals.

摘要

未标注

本研究旨在确定一种使用激光消融疗法(LA)联合吲哚菁绿(ICG)治疗感染根管的安全方案。

方法

体外研究——将拔除的人下颌切牙根管预备至#35/.06,并充满0.05%的ICG染料。在将#20光纤插入根管的过程中激活波长810 nm的红外二极管激光。根据方案,改变功率、脉冲间隔和脉冲持续时间,记录根管三个根段达到的最高温度。体内研究——选取拔除的人下颌切牙、下颌前磨牙和上颌磨牙。将其根管预备至#35/0.04,然后将牙齿植入Wistar大鼠的皮下组织。对每颗牙根进行体外研究中选定的LA联合ICG方案治疗。测量根管三个根段的表面温度。然后,立即或在7天后切除牙根周围的皮下组织,进行苏木精-伊红(H&E)染色的组织学分析。人体研究——纳入11例患者拟拔除的13颗下颌切牙。所有牙齿的根管均按与体内研究相似的方式进行预备并接受LA联合ICG治疗。拔牙后,对牙根根尖段进行组织病理学分析。

结果

体外研究表明,功率和脉冲持续时间的增加或脉冲间隔的缩短会导致温度升高。功率最高且温度升高低于40°C的方案为2.5 W(功率)、300 ms(间隔)和100 ms(持续时间)。体内研究表明,下颌切牙观察到的温度变化最大;因此,选择其进行人体研究。在人体研究中,在治疗后的牙齿中观察到牙周膜结构,其组织学和形态学特征与健康组织以及对照牙的牙根相似(p < 0.05)。在评估的任何牙齿中,未观察到对牙骨质结构、牙周膜和牙槽骨的有害影响。

结论

得出的结论是,所使用的LA方案是安全的,可在感染根管治疗期间进行。