Suppr超能文献

使用彩色多普勒超声对腹膜透析导管进行动态功能检查。体外技术评估及体内应用。

Dynamic functional examination of the peritoneal dialysis catheter using color Doppler ultrasound. In vitro technical evaluation and in vivo application.

作者信息

Zeiler Matthias, Mancini Anastasia, Federico Antonio, Ramazzotti Valentina, Silvestri Simona, Spatola Leonardo, Santarelli Stefano, Granata Antonio

机构信息

Nephrology and Dialysis Unit, Peritoneal Dialysis Section, Ospedale "C. E G. Mazzoni" Hospital, Via Degli Iris N. 1, 63100, Ascoli Piceno, Italy.

Nephrology and Dialysis Unit, "Carlo Urbani" Hospital, Jesi, Italy.

出版信息

J Nephrol. 2025 Sep 13. doi: 10.1007/s40620-025-02421-3.

Abstract

BACKGROUND

Peritoneal dialysis catheter malfunction compromises dialysis adequacy. Standard imaging methods in the case of peritoneal catheter malfunction include abdominal X-ray, fluoroscopic catheter peritoneography, and computed tomography. Ultrasound has only recently been utilized to evaluate the intraperitoneal part of the catheter. In clinical routine, catheter function is assessed by the time needed to fill or drain a defined quantity of peritoneal dialysis fluid. Visual functional testing of the catheter can be performed by fluoroscopy or by contrast-enhanced ultrasound. We developed and tested a color Doppler ultrasound technique for dynamic imaging of the peritoneal catheter based on signal generation by dialysis fluid flow.

METHODS

The feasibility of the color Doppler approach was evaluated utilizing a phantom. Furthermore, the technique was applied in 28 peritoneal dialysis patients, most of whom presented peritoneal catheter malfunction.

RESULTS

Color Doppler examination improved the catheter visualization, especially in cases in which the catheter is embedded between the intestinal loops. Furthermore, this technique highlighted catheter side hole occlusions by adhesions or intraluminal thrombi. The examination can be performed either during filling or draining of dialysis fluid.

CONCLUSIONS

The color Doppler ultrasound technique appears to be helpful in identifying problem-solving strategies in malfunctioning peritoneal catheters.

摘要

背景

腹膜透析导管功能障碍会影响透析充分性。腹膜导管功能障碍时的标准成像方法包括腹部X线、透视下导管腹膜造影和计算机断层扫描。超声直到最近才被用于评估导管的腹腔内部分。在临床常规中,通过填充或引流一定量腹膜透析液所需的时间来评估导管功能。导管的视觉功能测试可通过透视或对比增强超声进行。我们开发并测试了一种基于透析液流动产生信号的彩色多普勒超声技术,用于腹膜导管的动态成像。

方法

利用模拟体评估彩色多普勒方法的可行性。此外,该技术应用于28例腹膜透析患者,其中大多数存在腹膜导管功能障碍。

结果

彩色多普勒检查改善了导管的可视化,特别是在导管嵌入肠袢之间的情况下。此外,该技术突出显示了因粘连或管腔内血栓导致的导管侧孔闭塞。该检查可在透析液填充或引流期间进行。

结论

彩色多普勒超声技术似乎有助于确定处理功能障碍腹膜导管的解决策略。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验