Young I R, Bryant D J, Payne J A
Magn Reson Med. 1985 Aug;2(4):355-89. doi: 10.1002/mrm.1910020406.
Existing models for the contrast between one tissue and another in NMR imaging have shown that there are regions of ambiguity particularly in spin-echo sequences. This paper indicates that there are further problems due to variations in slice formation with tissue parameters, and that it is necessary to be very conservative in assessing the magnitude of T1 in any imaging system. The change in shape affects the appearance of images due, in essence, to adjustments in partial volume effects, with some of the contributions to the signal from lesions with long T1 coming from different regions to those for normal tissue. A simple model for a slice is used because the actual form in a real imaging situation is the result of a number of factors which are usually not accurately enough known.
核磁共振成像中用于区分一种组织与另一种组织的现有模型表明,尤其在自旋回波序列中存在模糊区域。本文指出,由于切片形成随组织参数的变化还存在其他问题,并且在评估任何成像系统中T1的大小时必须非常谨慎。形状的变化实际上会由于部分容积效应的调整而影响图像外观,T1较长的病变对信号的一些贡献来自与正常组织不同的区域。使用了一个简单的切片模型,因为实际成像情况下的实际形式是由许多通常了解得不够准确的因素造成的。