评估“真实的袋阀面罩辅助通气帮助”对改善院前环境中人工通气质量的作用:一项前后对照模拟人研究。

Evaluation of "Real BVM Help" for Improving Manual Ventilation Quality in the Prehospital Setting: A Before-After Manikin Study.

作者信息

Krammel Mario, Grassmann Daniel, Heinrich Lukas, Brock Roman, Kornfehl Andrea, Pagitz Nikolaus, Valentova Karolina, Veigl Christoph, Heider Sabine, Girsa Michael, Aigner Patrick, Hamp Thomas, Schnaubelt Sebastian

机构信息

Emergency Medical Service Vienna, Vienna, Austria.

PULS - Austrian Cardiac Arrest Awareness Association, Vienna, Austria.

出版信息

Open Access Emerg Med. 2025 Sep 11;17:257-265. doi: 10.2147/OAEM.S520921. eCollection 2025.

Abstract

BACKGROUND

Manual ventilation is a critical skill for emergency medical service (EMS) members. However, it is challenging in terms of correct ventilation rates and tidal volumes, with potentially severe adverse effects of hypo- and hyperventilation. Measuring the quality and involving real-time feedback may be effective in optimizing of manual ventilation.

METHODS

Data acquired retrospectively from a quality management project in 143 advanced emergency medical technicians were included. They performed bag ventilations on an intubated adult manikin for two minutes without any feedback system, and then another two minutes with the Real BVM Help device. Ventilation rates and volumes and their allocation in correct/recommended ranges were determined.

RESULTS

With the feedback device, correctly applied ventilation rates increased by 21% (63.6% in the correct range without vs 84.6% with the feedback device; p<0.001), and ventilation volumes improved by 41% (27% in the correct range without vs 68% with the feedback device; p<0.001). Without the device, the average ventilation rate was 10.5 ±3.1/minute, compared to 9.5 ±1.9/minute with the device. Ventilation volumes amounted to 370.6 ±84 mL without Real BVM Help, while when using it, 415.5 ±33.1 mL was noted.

CONCLUSION

Our data demonstrate significant improvements in ventilation rates and volumes when using a ventilation feedback device. This manikin study suggests a ventilation feedback device being beneficial for the use by EMS members, but our findings must be further validated in real-life conditions.

摘要

背景

人工通气是紧急医疗服务(EMS)人员的一项关键技能。然而,在正确的通气频率和潮气量方面具有挑战性,通气不足和通气过度可能会产生严重的不良影响。测量质量并提供实时反馈可能有助于优化人工通气。

方法

回顾性纳入了一项质量管理项目中143名高级急救医疗技术人员的数据。他们在没有任何反馈系统的情况下,对一名插管成年人体模型进行了两分钟的球囊通气,然后使用Real BVM Help设备又进行了两分钟的通气。确定了通气频率、潮气量及其在正确/推荐范围内的分布情况。

结果

使用反馈设备后,正确应用的通气频率提高了21%(无反馈时正确范围为63.6%,使用反馈设备时为84.6%;p<0.001),潮气量提高了41%(无反馈时正确范围为27%,使用反馈设备时为68%;p<0.001)。不使用该设备时,平均通气频率为10.5±3.1次/分钟,使用该设备时为9.5±1.9次/分钟。没有Real BVM Help时潮气量为370.6±84毫升,使用该设备时为415.5±33.1毫升。

结论

我们的数据表明,使用通气反馈设备时,通气频率和潮气量有显著改善。这项人体模型研究表明,通气反馈设备对EMS人员有益,但我们的研究结果必须在实际情况中进一步验证。