Nair R, Kartha M
AJR Am J Roentgenol. 1978 Jan;130(1):150-3. doi: 10.2214/ajr.130.1.150.
Dose distribution patterns of 4 MV x-rays from a Varian Clinac 4 linear accelerator are compared with those from a 60Co teletherapy machine. Field flatteners of lead and depleted uranium were used. Dose distribution maps display the off-axis high dose or "horns" with lead filters and circular dose distribution patterns. The uranium field flattener clearly "clipped the horns" and eliminated circular dose distribution patterns with larger fields.