Alanine aminotransferase and aspartate aminotransferase measurements with two automated analyzers, SMAC and the ABA-100, compared.

作者信息

Purcell G V, Behenna D B, Walsh P R

出版信息

Clin Chem. 1979 May;25(5):780-2.

PMID:436250
Abstract

Measurements of alanine aminotransferase and aspartate aminotransferase with the SMAC were evaluated for correlation with the ABA-100, precision, linearity, and carryover. We assayed 200 specimens with normal and abnormal aminotransferase activities with both the SMAC (y) and the ABA-100 (X). Linear regression analysis of the data yielded the following: alanine aminotransferase (r = 0.9732, y = 0.96x + 3.8); and aspartate aminotransferase (r = 0.9892, y = 0.90x + 2.1). Both aminotransferases demonstrated acceptable intra- and inter-assay variations with the SMAC and ABA-100. With the SMAC the upper limit of linearity for alanine aminotransferase was 350 U/L; that for asparate aminotransferase was 450 U/L. Carryover studies for SMAC indicate that specimens immediately following specimens with alanine aminotransferase activities greater 400 U/L and (or) aspartate aminotransferase activities greater than 500 U/L should be re-analyzed.