Wackers F J, Berger H J, Johnstone D E, Goldman L, Reduto L A, Langou R A, Gottschalk A, Zaret B L
Am J Cardiol. 1979 Jun;43(6):1159-66. doi: 10.1016/0002-9149(79)90148-6.
The intrinsic variability and accuracy of left ventricular ejection fraction determined by multiple gated cardiac blood pool imaging was evaluated in 83 patients. Ejection fraction by gated studies correlated well with data from first pass radionuclide angiocardiography (r = 0.94) and from contrast angiography (r = 0.84). Intra- and interobserver variabilities of absolute ejection fraction were minimal (mean +/- standard deviation 1.4 +/- 1.2 and 1.6 +/- 1.5 percent, respectively) and were not different for normal (ejection fraction 55 percent or greater) and abnormal patients. Ejection fraction was determined twice in 70 patients: on the same day at intervals separated by 1 to 2 hours (41 patients) and on 2 different days (29 patients). Ejection fraction ranged from 18 to 91 percent and was normal in 37 patients. There was no difference in mean serial variabilities of absolute ejection fraction for all repeat studies performed on the same and separate days (3.3 +/- 3.1 versus 4.3 +/- 3.1 percent (not significantly different). The mean variability of absolute ejection fraction for repeat studies in normal patients was significantly greater than in abnormal patients (5.4 +/- 4.4 versus 2.1 +/- 2.0 percent, P less than 0.01). The incidence rate of absolute interstudy changes of 5 percent or more was significantly higher in normal than in abnormal patients (P less than 0.01). This differential variability should be considered in interpreting sequential changes in left ventricular ejection fraction. To be attributed to nonrandom physiologic alterations, the absolute change in ejection fraction should be 10 percent or more in normal patients and 5 percent or more in abnormal patients.
我们对83例患者进行了评估,以确定多门控心血池显像测定左心室射血分数的内在变异性和准确性。门控研究得出的射血分数与首次通过放射性核素血管造影术(r = 0.94)及造影剂血管造影术(r = 0.84)的数据具有良好的相关性。绝对射血分数的观察者内和观察者间变异性极小(分别为平均±标准差1.4±1.2和1.6±1.5%),正常(射血分数55%或更高)和异常患者之间无差异。70例患者的射血分数测定了两次:41例患者在同一天间隔1至2小时进行,29例患者在不同的两天进行。射血分数范围为18%至91%,37例患者正常。在同一天和不同天进行的所有重复研究中,绝对射血分数的平均系列变异性无差异(3.3±3.1%对4.3±3.1%,无显著差异)。正常患者重复研究的绝对射血分数平均变异性显著大于异常患者(5.4±4.4%对2.1±2.0%,P<0.01)。正常患者绝对研究间变化5%或更多的发生率显著高于异常患者(P<0.01)。在解释左心室射血分数的连续变化时,应考虑这种差异变异性。要归因于非随机生理改变,正常患者射血分数的绝对变化应在10%或更多,异常患者应在5%或更多。