Suppr超能文献

Evaluation of the Beckman creatinine analyzer.

作者信息

Hicks J M, Iosefsohn M, Lewis S A

出版信息

Clin Chem. 1979 Jun;25(6):1005-8.

PMID:445800
Abstract

We compared results obtained with the Beckman Creatinine Analyzer (Beckman Instruments, Inc.) to those with the GEMSAEC Centrifugal Analyzer (Electro-Nucleonics, Inc.) and with a manual method for determination of creatinine in serum and urine. Linearity, stability, recovery, and precision were evaluated. Interference from bilirubin, protein, lipemia, uric acid, and hemolysis is negligible, but acetone and acetoacetate can interfere.