DePuey E G, Mathur V S, Garcia E G, Burdine J A
Radiology. 1979 Jul;132(1):161-7. doi: 10.1148/132.1.161.
A new mobile dual cardiac probe has been introduced for measurements of the left ventricular ejection fraction (LVEF). In 28 patients, correlation between LVEFs measured by the dual probe and biplane contrast ventriculography was 0.83. Central and peripheral injections yielded comparable curves, and the resultant LVEFs correlated well (R = 0.95). Correlation between LFEVs determined by the probe and by gated blood pool imaging in 43 patients was also 0.83. In 21 patients given three serial injections of 99mTc-albumin, the mean variation of the ejection fraction was 4.2 +/- 3.2%; however, reproducibility was unsatisfactory using 99mTc-sulfur colloid due to sequestration in hepatic background tissue.
一种用于测量左心室射血分数(LVEF)的新型移动式双心脏探头已被推出。在28例患者中,双探头测量的LVEF与双平面造影心室造影之间的相关性为0.83。中心注射和外周注射产生了可比的曲线,并且所得的LVEF相关性良好(R = 0.95)。在43例患者中,探头测定的LFEV与门控血池显像之间的相关性也为0.83。在21例接受三次连续注射99mTc-白蛋白的患者中,射血分数的平均变化为4.2 +/- 3.2%;然而,由于在肝脏背景组织中滞留,使用99mTc-硫胶体时再现性不令人满意。