Mann S K, Mann N S
Am J Gastroenterol. 1979 Jun;71(6):568-71.
The effects of electrocoagulation on the esophagus, stomach and duodenum of mongrel dogs were studied. A commercially available electrocoagulator unit and suction electrode were used. Various dial settings and duration times were used. The total number of sites electrocoagulated was 945; histologic section of each site was prepared. A dial setting of 7 and duration of 5 seconds was safe for the canine esophagus and duodenum; for the stomach it is a dial setting of 7 and duration of 3 seconds. This information will be useful in safe application of electrocoagulation to superficial bleeding lesions in the canine esophagus, stomach and duodenum.
研究了电凝对杂种犬食管、胃和十二指肠的影响。使用了市售的电凝器装置和吸引电极。采用了各种刻度盘设置和持续时间。电凝的部位总数为945个;对每个部位都制作了组织学切片。刻度盘设置为7、持续时间为5秒对犬的食管和十二指肠是安全的;对胃来说,刻度盘设置为7、持续时间为3秒是安全的。该信息将有助于在犬食管、胃和十二指肠的浅表出血性病变中安全应用电凝。