Influence of beam characteristics and detector size in negative-pion dose studies.

作者信息

Hamm R N, Wright H A, Turner J E

出版信息

Phys Med Biol. 1979 May;24(3):613-20. doi: 10.1088/0031-9155/24/3/012.

Abstract

Dose calculations were performed for a tissue phantom irradiated by uniform circular beams of negative pions with an assumed gaussian momentum distribution. The mean momentum of the pions was varied from 104.4 to 171.5 MeV/c (mean range 5-20 cm in unit-density tissue) and the momentum spread from 0 to 5% of the mean. Depth-dose curves are shown for different mean momenta and momentum spreads. The radial distribution of dose as a function of distance from the beam axis was computed at different depths for a beam with a mean momentum of 153.4 MeV/c and spread of 2%. The responses of detectors of different sizes used to measure centre-line dose for this beam were shown by calculating depth-dose curves for detectors of radii 0.5, 1.0, 1.5, 2.0, 3.0 and 4.0 cm. Calculations were also performed for beams having radii of 1, 2 and 3 cm. Depending on particular conditions, it appears that considerable care may often be needed to infer the absorbed dose at a given location in a phantom irradiated by a negative-pion beam.

摘要

针对由具有假定高斯动量分布的负π介子均匀圆形束照射的组织体模进行了剂量计算。π介子的平均动量在104.4至171.5 MeV/c之间变化(在单位密度组织中的平均射程为5 - 20 cm),动量展宽从0至平均动量的5%。给出了不同平均动量和动量展宽下的深度剂量曲线。对于平均动量为153.4 MeV/c且展宽为2%的束,计算了在不同深度处作为距束轴距离函数的剂量径向分布。通过计算半径为0.5、1.0、1.5、2.0、3.0和4.0 cm的探测器的深度剂量曲线,展示了用于测量该束中心线上剂量的不同尺寸探测器的响应。还对半径为1、2和3 cm的束进行了计算。根据特定条件,在由负π介子束照射的体模中推断给定位置处的吸收剂量时,似乎通常可能需要格外小心。