Jekowsky E, Trainer T D, Copeland B E
Am J Clin Pathol. 1979 Aug;72(2):146-50. doi: 10.1093/ajcp/72.2.146.
A competitive protein-binding method (CPB) for serum cortisol determination has been compared with a newly developed automated radioimmunoassay (RIA) technic with respect to reproducibility and correlation of patient results. The RIA method uses an immobilized anti-cortisol antibody covalently coupled to the inner surface of a plastic cartridge. A mixture of 3H-labeled and unlabeled cortisol is passed through the cartridge to achieve separation of free and bound cortisol. The bound cortisol is later eluted to regenerate the antibody for the next sample. The RIA procedure had a day-to-day coefficient of variation (CV) of 7% at three cortisol levels. The CPB method had a day-to-day CV of 10% at a concentration of 8 microgram/dl. A comparison of 67 patient samples showed no statistically significant difference between the methods. The regression equation was RIA = 0.96 CPB + 0.14 microgram/dl, with a correlation of 0.92.
已就血清皮质醇测定的竞争性蛋白结合法(CPB)与新开发的自动放射免疫测定(RIA)技术在重现性和患者结果的相关性方面进行了比较。RIA方法使用与塑料盒内表面共价偶联的固定化抗皮质醇抗体。将3H标记的皮质醇和未标记的皮质醇混合物通过该盒,以实现游离皮质醇和结合皮质醇的分离。随后洗脱结合的皮质醇,使抗体再生以用于下一个样本。RIA程序在三个皮质醇水平下的日间变异系数(CV)为7%。CPB方法在浓度为8微克/分升时的日间CV为10%。对67份患者样本的比较显示,两种方法之间无统计学显著差异。回归方程为RIA = 0.96 CPB + 0.14微克/分升,相关性为0.92。