Measurement of postoperative cardiac output by thermodilution in pediatric and adult patients.

作者信息

Moodie D S, Feldt R H, Kaye M P, Danielson G K, Pluth J, O'Fallon M

出版信息

J Thorac Cardiovasc Surg. 1979 Nov;78(5):796-8.

PMID:491734
Abstract

Serial cardiac output determinations were made by the thermodilution technique in 51 patients by means of a No. 2 Fr. thermistor catheter placed directly into the pulmonary artery at cardiac operation. Correlations were determined prospectively between thermodilution measurements of cardiac output and other commonly used indirect clinical parameters. Serial indicator dye-dilution curves were performed in 24 of these patients and compared with simultaneous thermodilution measurements. A high correlation (r = 0.97) was noted between dye curve measurements of cardiac output and thermal measurements. Statistically significant correlations were also seen between cardiac output and both the quality of the peripheral pulses and the duration of cardiopulmonary bypass, but no significant correlations were found between the measured cardiac outputs and other variables. This study confirms the necessity for direct measurement of cardiac output for its accurate assessment.

摘要

采用热稀释技术,通过将2号法国热敏电阻导管在心脏手术时直接插入肺动脉,对51例患者进行了连续心输出量测定。前瞻性地确定了心输出量的热稀释测量值与其他常用间接临床参数之间的相关性。对其中24例患者进行了连续指示剂稀释曲线测定,并与同时进行的热稀释测量结果进行了比较。发现心输出量的染料曲线测量值与热测量值之间存在高度相关性(r = 0.97)。心输出量与外周脉搏质量和体外循环持续时间之间也存在统计学上的显著相关性,但在测量的心输出量与其他变量之间未发现显著相关性。本研究证实了直接测量心输出量以进行准确评估的必要性。