Kalender W, Reither M, Schuster W
Pediatr Radiol. 1979 Oct;8(4):233-5. doi: 10.1007/BF00974041.
A simple dose-saving procedure for the diagnosis of congenital hip dysplasia employing a diaphragm in near-focus position is presented. Doses to the skin and the gonads are measured with a phantom, comparing the new and other techniques. With optimal techniques the skin dose is smaller than 2 mR and the gonadal doses lie around 0.1 mR for both male and female patients.
介绍了一种采用近焦位置的滤线栅进行先天性髋关节发育不良诊断的简单剂量节省程序。使用体模测量皮肤和性腺的剂量,比较新方法和其他技术。采用最佳技术时,男性和女性患者的皮肤剂量均小于2毫伦琴,性腺剂量约为0.1毫伦琴。