Amis A A, Hughes S, Miller J H, Wright V, Dowson D
Rheumatol Rehabil. 1979 Nov;18(4):230-4. doi: 10.1093/rheumatology/18.4.230.
Whilst developing an elbow endoprosthesis, the joint forces were estimated for patients with rheumatoid arthritis. Elbow flexion strength of rheumatoid patients was found to be 45% of normal. Muscle strengths and limb geometry data were found by a dissection technique, which allowed joint forces to be calculated during flexion, extension and abduction efforts. Forces up to 2.4 kN were predicted to act on the distal humerus, with similar forces acting in both radius and ulna. The skeletal structure is well adapted to carry the predicted forces, and onlay-type prostheses are recommended for elbow replacement.
在开发肘部假体时,对类风湿性关节炎患者的关节力进行了估算。发现类风湿患者的肘部屈曲力量为正常力量的45%。通过解剖技术获得了肌肉力量和肢体几何数据,这使得能够计算出屈曲、伸展和外展动作时的关节力。预计高达2.4 kN的力作用于肱骨远端,桡骨和尺骨上的作用力相似。骨骼结构能够很好地适应所预测的力,因此建议采用覆盖型假体进行肘部置换。