Ogan K, Katz E, Porro T J
J Chromatogr Sci. 1979 Nov;17(11):597-600. doi: 10.1093/chromsci/17.11.597.
The use of a cut-off filter has been compared with a monochromator for emission energy selection in the fluorescence detector used in the analysis of environmental samples for polycyclic aromatic hydrocarbons by liquid chromatography. The use of a cut-off filter can provide greater sensitivity for compounds that are well separated by the chromatography. However, samples that contain a large number of compounds often result in overlapping peaks, and for such samples, it has been found that the spectral selectivity of monochromatic emission selection reduces the interference due to compounds eluting near the compounds of analytical interest. This spectral selectivity results in greater sensitivity and a smaller quantitation error in the analysis of these complex samples.
在通过液相色谱法分析环境样品中的多环芳烃时,已将截止滤光片与单色仪用于荧光检测器中的发射能量选择进行了比较。对于色谱分离良好的化合物,使用截止滤光片可提供更高的灵敏度。然而,含有大量化合物的样品常常会导致峰重叠,对于此类样品,已发现单色发射选择的光谱选择性可减少因在分析感兴趣的化合物附近洗脱的化合物所产生的干扰。这种光谱选择性在分析这些复杂样品时可带来更高的灵敏度和更小的定量误差。