Abstract

Capabilities of the Redy cartridge for hemofiltrate regeneration were tested. Electrolytes, creatinine, BUN, uric acid, acid-base, glucose, heparin, fluoride and amino acids were measured in the cartridge inflow and outflow (V = 70 ml/min) over a four-hour period. There was complete adsorption of potassium, calcium, magnesium, creatinine, BUN, uric acid, phosphate and heparin. Sodium ions, hydrogen ions, fluoride ions and possibly other trace elements are released by the cartridge. Amino acid absorption is almost complete if the amino acids are aromatic, have two or more N-atoms or if they have S-atoms. When using hemofiltrate after sorbent regeneration for reinfusion to the patient, one has to consider electrolyte substitution (i.e., K, Ca, Mg), sodium balance, trace element metabolism, acid-base problems, amino-acid profiles and ammonia overload.

摘要

对Redy透析器进行了血液滤过液再生能力测试。在四小时内,以70毫升/分钟的流速对透析器进、出口处的电解质、肌酐、尿素氮、尿酸、酸碱值、葡萄糖、肝素、氟化物及氨基酸进行了测量。钾、钙、镁、肌酐、尿素氮、尿酸、磷酸盐及肝素被完全吸附。透析器释放出钠离子、氢离子、氟离子以及可能的其他微量元素。若氨基酸为芳香族、含有两个或更多氮原子或含有硫原子,则氨基酸的吸收几乎是完全的。当使用吸附剂再生后的血液滤过液回输至患者体内时,必须考虑电解质替代(如钾、钙、镁)、钠平衡、微量元素代谢、酸碱问题、氨基酸谱及氨过载。