Royal H D, Brown P H, Claunch B C
J Nucl Med. 1979 Sep;20(9):977-80.
The performances of two commerical gamma cameras, with crystals 1.3 and 0.6 cm thick, are compared. Phantom studies simulating clinical conditions showed no significant difference in performance at 140 keV. At 68 keV, the thinner crystal gave marginal improvements in camera performance with phantoms simulating clinical conditions. Frequent use with very low-energy emitters, such as Tl-201, would be needed to justify the expense required to refit the 1.3 cm camera with a 0.6 cm crystal.
比较了两台配备厚度分别为1.3厘米和0.6厘米晶体的商用伽马相机的性能。模拟临床情况的体模研究表明,在140 keV时性能无显著差异。在68 keV时,对于模拟临床情况的体模,较薄的晶体使相机性能略有改善。要证明将1.3厘米相机改装为配备0.6厘米晶体所需的费用合理,就需要经常使用极低能量发射体,如铊-201。