Coriell L L, McGarrity G J
Appl Microbiol. 1970 Sep;20(3):474-9. doi: 10.1128/am.20.3.474-479.1970.
In a horizontal back-to-front flow high-efficiency particulate air-filtered laminar hood, it is shown that a Blake bottle obstruction to the air flow causes a downstream cone of turbulent air which can draw microbial contamination into the work area of the hood. In controlled experiments, contamination with T3 coliphage was reduced by a series of perforations around the open edge of the hood which eliminates the cone of turbulent air. The average reduction in phage counts was 90.75, 86.79, 91.12, and 98.92%, depending upon the site of nebulization. The phage counts were reduced in 48 of the 51 tests.
在一台水平从前向后气流的高效空气过滤器层流罩中,研究表明,一个布莱克瓶对气流的阻碍会导致下游出现紊流空气锥,这会将微生物污染物吸入层流罩的工作区域。在对照实验中,通过在层流罩开口边缘周围设置一系列穿孔消除了紊流空气锥,从而减少了T3噬菌体的污染。根据雾化部位的不同,噬菌体计数的平均减少率分别为90.75%、86.79%、91.12%和98.92%。在51次测试中有48次噬菌体计数减少。