Heidt P J
Antonie Van Leeuwenhoek. 1982;48(4):373-81. doi: 10.1007/BF00418290.
A modified microbiological safety cabinet which can be used as a class II and a class III safety cabinet has been bacteriologically tested. This cabinet makes use of a high-speed down-flow air curtain in the front opening to minimize the amount of air escaping over the arms of the operator. By using artificial aerosols and a dummy or a test person placing his arms into the working opening of the cabinet, a transfer from the inside to the environment was detected only when the highest concentration of the test aerosol was used. Since the number of bacteria detected was very low, this is considered to be acceptable. When the cabinet was used as a class III type, with a glove panel mounted in the front opening, leakage from the environment occurred. This could be completely prevented by fixing tape over the hinge of the front panel. The conclusion is drawn that this type of biohazard hood can be safely used as a class II and a class III microbiological safety cabinet, provided the construction of the hinge of the front panel will be adapted to prevent transfer from the environment to the working area.