Suppr超能文献

一种用于监测儿童和成人硬膜外压力的新型小型化系统。

A new miniaturized system for monitoring the epidural pressure in children and adults.

作者信息

Dietrich K, Gaab M, Knoblich O E, Schupp J, Ott B

出版信息

Neuropadiatrie. 1977 Feb;8(1):21-8. doi: 10.1055/s-0028-1091501.

Abstract

A new miniaturized epidural pressure transducing system for monitoring intracranial pressure (ICP) has been developed. The system is of the as yet smallest known dimensions. The transducer is implanted automatically coplanar to the dura via a burr hole of 4 mm diameter in the conscious patient without special surgical equipment. Application is possible in adults, children and infants. We have achieved a maximal resolution of 10(-3) mm of Hg. At a full scale range of 100 mm of Hg the zero-point drift is less than 1 mm of Hg/24 h.

摘要

一种用于监测颅内压(ICP)的新型小型化硬膜外压力传感系统已被开发出来。该系统具有目前已知的最小尺寸。传感器通过直径4毫米的骨钻孔,在清醒患者身上自动与硬脑膜共面植入,无需特殊手术设备。该系统可应用于成人、儿童和婴儿。我们已实现10(-3)毫米汞柱的最大分辨率。在100毫米汞柱的满量程范围内,零点漂移小于1毫米汞柱/24小时。