Carlene E, Ericsson M, Levander-Lindgren M, Pehrsson K
Br Heart J. 1977 Nov;39(11):1265-72. doi: 10.1136/hrt.39.11.1265.
An atrial detector electrode was introduced by mediastinoscopy in 82 patients requiring permanent cardiac pacing. There were no complications. An adequate P wave was recorded in 80 patients. During the following week, the P wave became ineffective in 5 patients; angina occurred in 2 and atrial arrhythmias in 2. Atrially triggered ventricular pacing established in 73 patients and was followed in 71 patients for a period of 1 to 113 months. In 17 cases, it had to be terminated because of an ineffective or unstable P wave,in 6 cases because of atrial arrhythmias, and in 4 cases because of advanced age and recurrent infections. The method is technically simple and place little stress on the patient.
对82例需要永久性心脏起搏的患者,通过纵隔镜插入心房探测电极。无并发症发生。80例记录到合适的P波。在接下来的一周内,5例患者的P波变得无效;2例出现心绞痛,2例出现房性心律失常。73例患者建立了心房触发心室起搏,并对其中71例患者随访了1至113个月。17例因P波无效或不稳定而终止,6例因房性心律失常而终止,4例因高龄和反复感染而终止。该方法技术上简单,对患者压力小。