Improvements and simplifications of the ozonolysis-Schiff staining method for plasma lipoproteins on cellulose acetate.

作者信息

Uldall A

出版信息

J Clin Chem Clin Biochem. 1976 Jan;14(1):23-6. doi: 10.1515/cclm.1976.14.1-12.23.

Abstract

Modifications of the ozonolysis and Schiff staining of plasma lipoproteins reduce 11 pre-spectrometric steps to 3. Picrate denaturation of the lipoproteins and absorption spectrometry directly on the wet opaque membrane increase the sensitivity; the volume of sub-specimen needed per length of application site is reduced from 0.7 to 0.1 mul/mm. Reasonably good proportionality between the total lipoprotein concentration and the spectrometric findings are obtained. Two specimens were obtained at an interval of 8-13 days from each of 10 patients; for the arbitrary substance fraction of alpha-, prebeta-, and beta-lipoproteins in plasma the between-series intra-individual standard deviations were 0.021, 0.023 and 0.032 respectively.