Bacharach S L, Green M V, Borer J S, Ostrow H G, Redwood D R, Johnston G S
J Nucl Med. 1977 Dec;18(12):1176-83.
A nonimaging, ECG-gated scintillation-probe system is described that permits real-time quantification, at high temporal resolution, of the time variation of left ventricular (LV) volume over a complete, average cardiac cycle. Linearity between counting rate and volume, probe positioning, and background correction were investigated for both cylindrically collimated (CC) and parallel-hole-collimated (PC) detectors. In 53 patient studies, results obtained with these probes were compared with results obtained from an ECG-gated gamma camera system (CS) with high temporal resolution. Time-activity curves obtained by all three devices were essentially identical in shape (for CC against CS, r=0.93; for PC against CS, r=0.98) and in intracycle timing. Left-ventricular ejection fractions obtained with the probes showed workable agreement with the camera; for CC against CS, r=0.85 (N=31; for PC against CS, r=0.90 (N=21). When LV background is removed as a source of error, the correlation between (PC) probe and camera is improved (r=0.95, N=21). This suggests that the portable probe system be used in circumstances where exact knowledge of LV background is minimally important--e.g., continuous bedside monitoring of changes in LV function.
本文描述了一种非成像、心电图门控闪烁探头系统,该系统能够在高时间分辨率下实时定量分析整个平均心动周期内左心室(LV)容积随时间的变化。针对圆柱形准直(CC)探测器和平行孔准直(PC)探测器,研究了计数率与容积之间的线性关系、探头定位以及背景校正。在53例患者研究中,将这些探头获得的结果与通过具有高时间分辨率的心电图门控伽马相机系统(CS)获得的结果进行了比较。所有三种设备获得的时间-活性曲线在形状(CC与CS相比,r=0.93;PC与CS相比,r=0.98)和心动周期内定时方面基本相同。用探头获得的左心室射血分数与相机显示出可行的一致性;CC与CS相比,r=0.85(N=31);PC与CS相比,r=0.90(N=21)。当去除左心室背景作为误差源时,(PC)探头与相机之间的相关性得到改善(r=0.95,N=21)。这表明便携式探头系统可用于对左心室背景的确切了解不太重要的情况,例如对左心室功能变化进行连续床边监测。