Den Dulk K, Lindemans F W, Bär F W, Wellens H J
Pacing Clin Electrophysiol. 1982 Jul;5(4):476-85. doi: 10.1111/j.1540-8159.1982.tb02264.x.
Three cases of pacemaker interactive tachycardia are presented. The first two are [artificial] circus movement tachycardias. In the first one the retrograde arm of the tachycardia circuit was provided by the A-V node and the antegrade arm by an atrial synchronous pulse generator. In the second case, the A-V node and, coincidentally, an A-V sequential pulse generator alternately provided the antegrade arm while the retrograde arm was by way of an accessory pathway. In the third case ventricular inhibition during A-V sequential pacing gave the paced atrial events the chance to be conducted to the ventricles with a long A-V interval. This resulted in a tachycardia with a rate of 150 bpm, instead of the programmed rate of 110 bpm.
本文报告了3例起搏器介导性心动过速。前两例为[人工]折返性心动过速。第一例中,心动过速环路的逆行支由房室结提供,顺行支由心房同步起搏器提供。第二例中,房室结以及巧合的是,房室顺序起搏器交替提供顺行支,而逆行支通过旁路传导。第三例中,房室顺序起搏时的心室抑制使起搏的心房激动有机会以较长的房室间期下传至心室。这导致了心率为150次/分的心动过速,而不是程控的110次/分。