Scheibelhofer W, Glogar D, Probst P, Milczoch J, Kaindl F
Pacing Clin Electrophysiol. 1982 Jul;5(4):587-92. doi: 10.1111/j.1540-8159.1982.tb02284.x.
The induction of ventricular tachycardia or ventricular fibrillation by competitive pacing, especially in the setting of acute myocardial ischemia, is well known. A case of ventricular tachycardia induced by a multiprogrammable unipolar cathodal ventricular pacemaker is reported. The arrhythmia was caused by reprogramming, which necessitates a short switch to fixed rate pacing in this model (Spectrax 5985 SX). This potential hazard is not well established in patients with unipolar pacemakers. The use of the magnet as one of the preconditions for reprogramming should be avoided in future pacemakers.
竞争性起搏诱发室性心动过速或心室颤动,尤其是在急性心肌缺血的情况下,这是众所周知的。本文报告了一例由多程控单极阴极心室起搏器诱发室性心动过速的病例。心律失常是由重新编程引起的,在该型号(Spectrax 5985 SX)中,这需要短暂切换到固定频率起搏。这种潜在风险在单极起搏器患者中尚未得到充分证实。未来的起搏器应避免将使用磁铁作为重新编程的前提条件之一。