Mintzer R A, Matthies H J, Lin P P, Neiman H L, Rogers L F
Surg Gynecol Obstet. 1978 Jan;146(1):79-80.
The Min-R system is more than a satisfactory substitute for xerography. At equal distances, the images obtained are equivalent. Radiation exposure was 25 per cent of xerography, a significant reduction in exposure. By using a long cone and improving geometry, superior images were obtained in 4 per cent of 400 comparison patients. If mammography is to be used as an early detection device in patients with carcinoma of the breast, modalities with lower exposure requirements must be used. Min-R, a new film screen system, fulfills the requirement of reduced patient exposure without loss of diagnostic accuracy.
Min-R系统是一种比静电复印术更令人满意的替代方法。在相同距离下,所获得的图像是等效的。辐射暴露量仅为静电复印术的25%,显著降低了暴露量。通过使用长锥形并改进几何形状,在400例对照患者中有4%获得了优质图像。如果乳腺摄影要用作乳腺癌患者的早期检测手段,就必须使用对暴露要求较低的方式。Min-R,一种新型的胶片屏系统,满足了减少患者暴露量且不损失诊断准确性的要求。