Suppr超能文献

Evaluation of the Harleco Micro CO2 system for measurement of total CO2 in serum or plasma.

作者信息

Lam C W, Tan I K

出版信息

Clin Chem. 1978 Jan;24(1):143-5.

PMID:618644
Abstract

In the Harleco Micro CO2 System a volumetric procedure is used to measure total CO2 content in a small volume (0.10 ml) of serum or plasma. CO2 gas is released by mixing specimen and acid, either manually or with a "Micro CO2 Shaker." Within-batch CV's by manual and machine shakings were, respectively, 7.7 and 5.6% for a total CO2 value of 8 mmol/liter, decreasing to 2.0 and 1.4% for a value of 45 mmol/liter. Between-batch CV's were 8.0 and 7.1% at low total CO2, decreasing to 2.4 and 2.0% at high total CO2. Measurements were linear for values from 5 to 60 mmol/liter. Analytical recoveries at low and high concentrations were 97.7 to 98.9%. Correlation coefficients for results obtained with the Technicon "SMA Plus" autoAnalyzer and the Radiometer ABL 1 blood gas analyzer were 0.992 and 0.996, respectively. The Harleco Micro CO2 System appears to be a quick, reliable, and inexpensive method for measuring serum or plasma total CO2.