Forty-nine clinical isolates of group B streptococci (GBS) were correctly grouped from broth culture by the Fluoro-Kit immunofluorescence test. A further 82 beta-haemolytic streptococci of groups A, C, D, F, and G were tested and gave no cross-reactions. The test was simple to perform and gave clear results. The Fluoro-Kit reagents, however, failed to detect GBS in 21 (51%) of 41 smears of rectal or vaginal swabs from pregnant women from which GBS were subsequently grown. Thirty-two (20%) of 159 culture-negative swabs gave positive immunofluorescent reactions.
通过Fluoro-Kit免疫荧光试验,从肉汤培养物中正确鉴定出49株B族链球菌(GBS)临床分离株。对另外82株A、C、D、F和G组的β溶血性链球菌进行检测,未出现交叉反应。该试验操作简单,结果清晰。然而,在41份来自孕妇的直肠或阴道拭子涂片(随后从中培养出GBS)中,Fluoro-Kit试剂未能检测出21份(51%)中的GBS。在159份培养阴性的拭子中,有32份(20%)出现了阳性免疫荧光反应。