[Digital subtraction angiocardiography: accuracy of measuring left ventricular wall thickness and wall volume during intravenous contrast media injection].

作者信息

Lauber A, Deetjen W, Jehle J, Pölitz B, Schmiel F K, Spiller P

出版信息

Z Kardiol. 1984 Apr;73(4):257-63.

PMID:6375186
Abstract

A comparison of left ventricular wall thickness and wall volume determined by intravenous digital subtraction angiocardiography and by conventional left ventricular angiocardiography was performed by linear regression analysis: wall thickness: WDDSA = 0.90 X WDLV , Syx = +/- 0.16 cm, r = 0.74; wall volume: WVOLDSA = 0.87 X WVOLLV , Syx = +/- 48 ml, r = 0.85. Compared to the deviation and scatter of corresponding data measured from two consecutive beats in conventional left ventricular angiocardiographies , the error of determination of wall thickness and wall volume, determined by intravenous digital subtraction angiocardiography, is more pronounced. The regression analysis of wall volumes under rest and bicycle exercise evaluated by digital subtraction angiocardiography revealed a similar inaccuracy ( Syx = +/- 46 ml). This is due to faults in recognition of the inner and outer contours of left ventricular wall. Our results show that measurements of left ventricular wall thickness and wall volumes by means of intravenous digital subtraction angiocardiography can only be performed with greater deviations compared to conventional left ventricular angiocardiograms.

摘要

通过线性回归分析比较了经静脉数字减影血管造影术和传统左心室血管造影术测定的左心室壁厚度和壁容积

壁厚度:WDDSA = 0.90×WDLV,Syx = ±0.16 cm,r = 0.74;壁容积:WVOLDSA = 0.87×WVOLLV,Syx = ±48 ml,r = 0.85。与传统左心室血管造影术中连续两个心动周期测量的相应数据的偏差和离散度相比,经静脉数字减影血管造影术测定的壁厚度和壁容积的误差更为明显。经数字减影血管造影术评估的静息和自行车运动状态下壁容积的回归分析显示出类似的不准确性(Syx = ±46 ml)。这是由于左心室壁内、外轮廓识别错误所致。我们的结果表明,与传统左心室血管造影相比,经静脉数字减影血管造影术测量左心室壁厚度和壁容积时偏差更大。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验