Colebatch H J, Nail B S, Ng C K
J Appl Physiol Respir Environ Exerc Physiol. 1978 Apr;44(4):611-8. doi: 10.1152/jappl.1978.44.4.611.
A system devloped for on-line measurement of transpulmonary pressure, gas flow at the mouth, change in expired volume and plethysmograph volume uses a minicomputer to control a multiplexed analog to digital converter. The computer identified samples as static or dynamic values by monitoring a voltage activating a solenoid valve, used to close the airway. Analysis of these samples by other task-specific programs yielded the static deflation pressure-volume (PV) curve, the conductance-recoil pressure, GL-Pst(L), relationship and the maximum expiratory flow-volume (MEFV) curve; the MEF-Pst(L) curve and conductance upstream from the equal pressure point were derived. The PV relationship was represented by a fourth-order polynomial and the GL-Pst(L) relationship by linear regression. In 11 subjects the results obtained using on-line data collection, compared with manual analysis of oscillograph recordings, showed small differences in static compliance and in the maximum Pst(L); but overall the two methods showed excellent agreement. Besides advantages of speed and objectivity, this system facilitates a more rigorous analytical treatment of elastic recoil and conductance.
一个用于在线测量跨肺压、口腔气体流量、呼出气体量变化和体积描记器体积的系统,使用一台小型计算机来控制一个多路复用模数转换器。计算机通过监测一个用于关闭气道的电磁阀的激活电压,将样本识别为静态或动态值。由其他特定任务程序对这些样本进行分析,得出静态放气压力-容积(PV)曲线、传导率-弹性回缩压力(GL-Pst(L))关系以及最大呼气流量-容积(MEFV)曲线;还得出了MEF-Pst(L)曲线以及等压点上游的传导率。PV关系用四阶多项式表示,GL-Pst(L)关系用线性回归表示。在11名受试者中,与对示波器记录进行手动分析相比,使用在线数据收集获得的结果显示,静态顺应性和最大Pst(L)存在微小差异;但总体而言,这两种方法显示出极好的一致性。除了速度和客观性方面的优势外,该系统有助于对弹性回缩和传导率进行更严格的分析处理。