Abstract

A circuit was designed for integrating respiratory flow signals over time to produce volume signals. The circuit involves a simple integrator and three compensation subcircuits which control integrator drift caused by ambient temperature changes, flow signal base-line errors, and random signal errors. A functional unit built according to the circuit design requires initial adjustment, but no further adjustment before or during sustained use. Integration errors are typically less than 1%. The only flow signal requirements are that they be bidirectional and within the human physiological frequency range.

摘要

设计了一种电路,用于对呼吸流量信号进行随时间的积分以产生容积信号。该电路包括一个简单的积分器和三个补偿子电路,这三个补偿子电路可控制由环境温度变化、流量信号基线误差和随机信号误差引起的积分器漂移。根据该电路设计构建的功能单元需要进行初始调整,但在持续使用之前或期间无需进一步调整。积分误差通常小于1%。对流量信号的唯一要求是它们必须是双向的且在人体生理频率范围内。