热式热电偶流量计的一个严重问题。

A serious problem of the heated thermocouple flowmeter.

作者信息

Satoh M, Kumada K, Tsunekawa K, Soma T

出版信息

Jpn Circ J. 1980 Sep;44(9):762-9. doi: 10.1253/jcj.44.762.

Abstract

In the measurement of regional blood flow with heated thermocouple flowmeter, we frequently observed puzzling phenomenon that the measured values were augmented while the flow was actually diminished and vice versa (Inversion). "Inversion" was directly attrebutable to the structure of the thermocouple probe. A countermeasure against "Inversion" was found to lie in the arrangement of the hot junction in the center of the heater, and in practical application, it must be checked by stopping the flow at each measurement. Overcoming this technical liability should pave the way for a marked improvement in the reliability of this method which, despite possessing many excellent characteristics for determining regional blood flow, is not widely used.

摘要

在用加热式热电偶流量计测量局部血流时,我们经常观察到一种令人困惑的现象,即实际血流减少时测量值却增大,反之亦然(反转)。“反转”直接归因于热电偶探头的结构。发现防止“反转”的对策在于将热接点布置在加热器的中心,并且在实际应用中,每次测量时必须通过停止血流来进行检查。克服这一技术缺陷应为显著提高该方法的可靠性铺平道路,尽管该方法在测定局部血流方面具有许多优异特性,但尚未得到广泛应用。