Banis J C, Schwartz K S, Acland R D
Plast Reconstr Surg. 1980 Oct;66(4):534-44. doi: 10.1097/00006534-198010000-00006.
We describe a reliable experimental method for direct, continuous measurement of the rate of blood flow in an island skin flap, using an electromagnetic flowmeter applied to the artery of the flap. The canine saphenous island flap model, developed as part of this study, is a large (11 x 14 cm) island flap, based solely on the saphenous artery (2 mm in diameter). We describe the anatomy of the flap and the method of raising it. Electromagnetic flowmetry is the only method of blood flow determination that provides immediate, continuous, and quantitative measurement of flow. We describe the principles and pitfalls of the electromagnetic flowmeter and the numerous rules of practice that must be observed to obtain consistent results. A validation study was carried out, in which we simultaneously measured arterial inflow with the flowmeter and venous outflow from the flap using direct collection. This study was based on the assumption that at any point in time the arterial and venous rate of flow were equal. We present the results of the study, which show an extremely close linear relationship between the measured and actual rates of flow. We intend to use this experimental method to study the factors that affect the rate of blood flow in free skin flap transfer.
我们描述了一种可靠的实验方法,该方法使用应用于皮瓣动脉的电磁流量计,直接、连续地测量岛状皮瓣中的血流速率。作为本研究一部分开发的犬隐静脉岛状皮瓣模型是一个大的(11×14厘米)岛状皮瓣,仅基于隐动脉(直径2毫米)。我们描述了皮瓣的解剖结构及其掀起方法。电磁血流测量法是唯一能即时、连续且定量测量血流的血流测定方法。我们描述了电磁流量计的原理和陷阱以及为获得一致结果必须遵守的众多操作规则。进行了一项验证研究,在该研究中,我们使用流量计同时测量动脉流入量,并通过直接收集测量皮瓣的静脉流出量。本研究基于这样的假设:在任何时间点,动脉和静脉的血流速率相等。我们展示了该研究的结果,结果表明测量的血流速率与实际血流速率之间存在极其紧密的线性关系。我们打算使用这种实验方法来研究影响游离皮瓣移植中血流速率的因素。