Marple V A, Rubow K L
Am Ind Hyg Assoc J. 1978 Mar;39(3):210-8. doi: 10.1080/0002889778507744.
A portable battery-operated optical particle counter/multichannel analyzer system has been developed for the numbers size distribution and number concentration measurement of light-absorbing irregular-shaped dust particles. An inertial impactor technique has been used to obtain calibration curves by relating the magnitude of the optical counter's signal to the particle's aerodynamic or Stokes' diameter. These calibrations have been made for aerosols of coal, potash, silica, rock (copper ore), and Arizona road dust particles.
已开发出一种便携式电池供电的光学粒子计数器/多通道分析仪系统,用于测量吸光不规则形状尘埃颗粒的数量粒径分布和数浓度。采用惯性撞击器技术,通过将光学计数器信号的大小与颗粒的空气动力学直径或斯托克斯直径相关联来获得校准曲线。已针对煤、钾盐、二氧化硅、岩石(铜矿石)和气溶胶以及亚利桑那道路尘埃颗粒进行了这些校准。