Benitez J T, Bouchard K R, Choe Y K
Ann Otol Rhinol Laryngol. 1978 Mar-Apr;87(2 Pt 1):216-23. doi: 10.1177/000348947808700212.
A major problem with air calorics appears to be one of technique. A pilot study led to the design of irrigating tips which allow consistant air presentation and simultaneous measurements of irrigating temperature at the delivery orifice (adding a second sensor). Preset temperatures of 24 C and 50 C in our system yielded air equilibration mean temperatures of 27.4 C and 45 C at the delivery orifice during testing. A normal study was carried out at these temperatures with an air flow of six liters per minute for 60 seconds. The range of caloric responses, mean maximum speed, and standard deviation are comparable to values previously reported with water stimulations. Statistical testing indicated no difference between ears for either warm or cold. Also, there was no significant difference for warm versus cold responses. We have performed over 2000 clinical examinations that incorporate this technique with satisfactory results and remarkable acceptance by the patients. The normal or "standard" probe size has been found to be adequate for the majority of the clinic population.
空气热量计的一个主要问题似乎是技术问题。一项初步研究促成了灌溉探头的设计,该探头能够实现稳定的空气输送,并能在输送口同时测量灌溉温度(增加了第二个传感器)。在我们的系统中,预设温度为24摄氏度和50摄氏度时,测试期间输送口处空气平衡的平均温度分别为27.4摄氏度和45摄氏度。在这些温度下进行了一项常规研究,空气流速为每分钟6升,持续60秒。热量反应范围、平均最大速度和标准差与先前用水刺激报告的值相当。统计测试表明,无论是温热还是冷刺激,双耳之间均无差异。此外,温热与冷刺激反应之间也没有显著差异。我们已经进行了2000多次采用该技术的临床检查,结果令人满意,患者接受度也很高。已发现正常或“标准”探头尺寸适用于大多数临床人群。